4530 Precision RTD simulator

The simulator is suitable for calibration laboratory and service-centers which need a computer-controlled resistance decade for the automation of calibration procedures.

Product Inquiry

    Product comparison

      • Features

        • Simulation of Pt100, Pt200, Pt500, Pt1000, Ni100 and Ni1000 sensors
        • Accuracy 0.02 °C
        • Temperature scales ITS 90, IPTS 68
        • Real Ohm simulation
        • Control manually or remotely via RS232/IEEE488/USB Converter
      • Product data

        Model 4530
        Pt100 simulation range

        -200.000 ... 850.000 °C

        Measurement accuracy 0.02 °C
        Pt200, Pt500, Pt1000 simulation Yes
        Ni100, Ni1000 simulation Yes
        Supply voltage 100 … 240 V, 50 / 60 Hz
        Interfaces

        • RS232
        • IEEE485
        • USB
        Specific characteristics

        • Temperature scales ITS 90, IPTS 68
        • Real Ohm simulation
        • Control manually or remotely via RS232/IEEE488/USB Converter
      • Description

        The precision RTD simulator is used where measuring instruments or controlling means with high precision must be examined or calibrated. The simulator is suitable for calibration laboratory and service centers which need a computer-controlled resistance decade for the automation of calibration procedures.

        All standard platinum and nickel RTD can be simulated with very high accuracy according to scales ITS 90 or IPTS 68.

        Temperature regulators, transducers etc. often supply a pulsed signal current in order to reduce the self-heating of the RTD. Devices with electronic simulation of RTD are therefore inclined to swinging and do not function with this application. Here, the advantage of the RTD simulator model 4530 becomes obvious as it can simulate the resistance sensor by real Ohm simulation without problems.

        Of course the RTD simulator can be used also as classical resistance decade.

        The equipment comes in a sturdy aluminium housing. The keyboard and display are on the front plate. It makes 2-, 3- and 4-wire connection possible at the back side.

        Pt100, Pt200, Pt500, Pt1000, Ni100 or Ni1000 can be selected by simple menu navigation. In addition, the selection of temperature scales ITS90 (DIN EN 60751) or IPTS 68 (DIN 43760) and two coefficients α = 0.385 (European) or α = 0.392 (US) is possible. After entry of the characteristic values the input of the temperature levels can be made very simply by the numeric keyboard. Presently set values are indicated in the display.

        Relays with low thermovoltage and stable foil resistances with very small temperature coefficients are the heart of the simulator.

      • Data Sheet & Downloads

        Data sheet
        Operation manual
        Software


      Brandnew Brochure: Load cells

      Discover our new brochure Load cells for production, automation, R&D and quality assurance with many application examples and product focused information!

      Download our new brochure

      burster in Malta

      Malta

      burster Italia S.r.l.

      Via Repubblica, 10
      24035 Curno
      Italy

      Fon: +309 035 618120
      Fax: +39 035618250
      infoburster.it
      www.burster.it

      SERVICES

      • Qualified design, application support
      • Process optimization
      • Qualified integration support & training
      • On-site trouble shooting
      • Qualified phone & online support
      • Local calibration service
      ×
      © 2024 // burster Italia s.r.l.
      nach oben