High-speed measurements from 10 ms/measurement, including evaluation
Temperature compensation for all materials
Thermoelectric voltage compensation
Input protection up to 400 Veff
32 adjustable measuring programs
Dry circuit measurement in accordance with DIN IEC 512
Product data
Model
2311
Measuring range
20.000 mΩ
200.00 mΩ
2.0000 Ω
20.000 Ω
200.00 Ω
2.0000 kΩ
20.000 kΩ
200.00 kΩ
Resolution
1 μΩ
10 μΩ
100 μΩ
1 mΩ
10 mΩ
100 mΩ
1 Ω
10 Ω
Measurement accuracy
0.03 % of reading ±2 digits
Automatic temperature compensation
10 different temperature coefficients can be selected and individually set
Interfaces
EtherCat
PROFINET
EtherNet/IP
Specific characteristics
Measuring ranges of 20 mΩ … to 200 kΩ
Resolution up to 1 μΩ
Measurement accuracy ≤ 0.03 % of reading
High-speed measurements from 10 ms/measurement, including evaluation
Temperature compensation for all materials
Thermoelectric voltage compensation
Input protection up to 400 Veff
32 adjustable measuring programs
Dry circuit measurement in accordance with DIN IEC 512
Description
The RESISTOMAT® model 2311 has been designed and optimized for high-speed applications in automation systems. Up to 100 measurements per second can be achieved. It works on the basis of the welltried four-wire measurement method in which test-lead resistances and contact resistances are eliminated. The instrument leads are monitored for damage by a built-in open circuit detector.
A 2-way and 4-way comparator with switching outputs is available for classifications and selections. Of course, temperature compensation is available for any test object material. Specific temperature coefficients can be entered. Temperature recording takes place using a PT100 sensor or a temperature transmitter (pyrometer) with an analog output.
A special circuit for protecting the measurement input when measuring inductive test objects has been developed to prevent damage to the meter from voltage peaks produced when the test object is disconnected.
A special area of application is the measuring of contact resistances (dry circuit measurement), since the load voltage is limited to 20 mV in order to avoid so-called “fritting” (DIN IEC 512).
All device settings can be individually stored in up to 32 measuring programs. Of course, all device settings can also be made via the Ethernet, USB (default) or fieldbus interfaces (optional). Up to 900 measurements per measuring program can be stored using the integrated data logger.
Discover our new brochure Load cells for production, automation, R&D and quality assurance with many application examples and product focused information!
The measurement solution. It‘s with this ambition and for over half a century burster, a global player in the precision measurement technology industry, supplies some of the world’s most recognizable brands with sensors, precision measuring devices and measuring systems for sensor signaling processing. This is all done through our research, development and production headquarters in Germany.